IEC 60567 pdf download

admin
IEC 60567 pdf download

IEC 60567 pdf download.Oil-filled electrical equipment – Sampling of gases and analysis of free and dissolved gases – Guidance
1 Scope
This International Standard deals with the techniques for sampling free gases from gas- collecting relays from power transformers. Three methods of sampling free gases are described. The techniques for sampling oil from oil-filled equipment such as power and instrument transformers, reactors, bushings, oil-filled cables and oil-filled tank-type capacitors are no longer covered by this standard, but are instead described in 4.2 of IEC 60475:201 1 . Before analysing the gases dissolved in oil, they are first extracted from the oil. Three basic methods are described, one using extraction by vacuum (Toepler and partial degassing), another by displacement of the dissolved gases by bubbling the carrier gas through the oil sample (stripping) and the last one by partition of gases between the oil sample and a small volume of the carrier gas (headspace). The gases are analysed quantitatively after extraction by gas chromatography; a method of analysis is described. Free gases from gas-collecting relays are analysed without preliminary treatment. The preferred method for assuring the performance of the gas extraction and analysis equipment, considered together as a single system, is to degas samples of oil prepared in the laboratory and containing known concentrations of gases (“gas-in-oil standards”) and quantitatively analyse the gases extracted. Two methods of preparing gas-in-oil standards are described. For daily calibration checks of the chromatograph, it is convenient to use a standard gas mixture containing a suitable known amount of each of the gas components to be in a similar ratio to the common ratios of the gases extracted from transformer oils . The techniques described take account, on the one hand, of the problems peculiar to analyses associated with acceptance testing in the factory, where gas contents of oil are generally very low and, on the other hand, of the problems imposed by monitoring equipment in the field.
3 Sampling of gases from gas-collecting (Buchholz) relays
3.1 General remarks It is important to bear in mind that receiving a qualitative and a representative sample is crucial for obtaining a reliable diagnosis of the electrical equipment. Even the most sophisticated extraction or diagnosis methods cannot overcome faulty samples. Gas samples from relays should be taken from the equipment with the minimum delay after gas accumulation has been signalled. Changes in composition caused by the selective re- absorption of components may occur if free gases are left in contact with oil. Certain precautions are necessary when taking gas samples. The connection between the sampling device and the sampling vessel shall avoid the ingress of air. Temporary connections should be as short as possible. Any rubber or plastic tubing used should have been proved to be impermeable to gases. Gas samples should be properly labelled (see Clause 4) and analysed without undue delay to minimize hydrogen loss (for example, within a maximum period of one week). Oxygen, if present in the gas, may react with any oil drawn out with the sample. Reaction is delayed by excluding light from the sample, for example, by wrapping the vessel in aluminium foil or suitable opaque material. Of the three methods described below, the syringe method is recommended. The other two methods are alternatives to be used exclusively in case of serious hindrance. Sampling into a sampling tube by liquid displacement using transformer oil as a sealing liquid is simple, but the different solubilities of the gas components may need to be taken into account if the gas quantity is such that some oil remains in the tube. The vacuum method requires skill to avoid contaminating the sample by leakage of air into the system. It is particularly true where the gas to be sampled may be at less than atmospheric pressure (for example, some sealed transformers).